知镜 Cognispect · 医学文献与研究工作台

Diabetic ketoacidosis associated with the reuse of single-use materials in a patient with type 1 diabetes mellitus treated with an advanced hybrid closed-loop system: A case report.

López Calles A, Plaza Chaparro P, Moreno Fernández J.
Endocrinologia, diabetes y nutricion · 2026-11-07 · PMID 42998935

文献摘要

Diabetic ketoacidosis (DKA) remains a potentially serious complication in individuals with type 1 diabetes mellitus, even in those treated with continuous subcutaneous insulin infusion systems. We describe the case of a 45-year-old woman using an advanced hybrid closed-loop system who was admitted with an episode of mild DKA in the absence of identifiable precipitating factors. Metabolic correction was performed according to current recommendations, with a favorable clinical course. Subsequent review of continuous glucose monitoring data and insulin delivery records revealed persistent hyperglycemia secondary to malfunction of the infusion device. This was associated with structural deterioration of a reused insulin cartridge designed for single use, in the context of a shortage of original supplies. This case highlights the importance of infusion materials in pump therapy safety, the limitations of automated systems in detecting mechanical failures, and the need for safe alternative strategies in situations of supply disruption.

在知镜精读原文来源DOI 原文

知镜提供医学文献检索、阅读笔记、人体图谱、统计绘图与研究复现工具。