Diagnostic Accuracy of the BlueDop Vascular Expert for Detecting Peripheral Arterial Disease in People with Suspected Intermittent Claudication Referred to Secondary Care: A Cross-Sectional Study.
文献摘要
Background Manual ankle-brachial pressure index (ABI) assessment reqiuires significant time and expertise, precluding its use in primary care. Automated vascular assessment tools may offer a simple and efficient alternative for detecting peripheral arterial disease (PAD). However, further evidence is needed before they can be recommended. This study considers the diagnostic accuracy of the BlueDop vascular expert (BVE) in people with suspected intermittent claudication. Methods We performed a cross-sectional cohort study including people referred to our nurse-led intermittent claudication clinic. BVE measurements were compared to the manual resting ABI/toe brachial index (TBI) method, using traditional cutoff values of Results 217 participants and 430 limbs were included, with an overall PAD prevalence of 46%. Sensitivity, specificity, positive predictor values (PPVs) and negative predictor values (NPVs) for the BVE were 88.5%, 80.7%, 81.3% and 88.1% respectively, when compared to the manual ABI. When compared to the manual TBI, the sensitivity, specificity, PPV and NPV for the BVE were 45.5%, 88%, 62.5%, and 78.6% respectively, though this analysis included only 38 limbs. No BVE diagnoses were based on ABI alone, 71% were based on waveform alone and 29% were based on a combination of ABI and waveform. The BVE assessment was significantly quicker to perform than the manual method (P Conclusion The BVE demonstrated favorable sensitivity and specificity for detecting or excluding PAD compared to the manual ABI. Further work is needed comparing this device to others, across a range of clinical populations and settings before it can be considered in routine clinical practice.