知镜 Cognispect · 医学文献与研究工作台

Diagnostic Accuracy and Operational Feasibility of the MESI mTABLET in Secondary-Care Diabetic Foot Clinics: A Pilot Cross-Sectional Service Evaluation.

Muscogliati R, Muscogliati E, Hitchman L, Lathan R, Smith G, Carradice D, Chetter I, Pymer S.
Annals of vascular surgery · 2026-09-18 · PMID 42759935

文献摘要

Background Peripheral arterial disease (PAD) is a common and serious complication in people with diabetes, particularly those with diabetic foot disease. Although measurement of the ankle-brachial pressure index (ABPI) and/or toe-brachial pressure index (TBPI) are first-line investigations for PAD, they are time-consuming and require training. Automated devices such as the MESI mTABLET may overcome these barriers, but diagnostic accuracy in diabetic populations is limited. Methods We conducted a cross-sectional diagnostic accuracy study in patients attending joint vascular and podiatry diabetic foot clinics at a UK tertiary center. Participants requiring ABPI assessment were considered for inclusion. Manual ABPI/TBPI measurements were obtained according to guidelines and served as the reference standard. Automated measurements were taken using the MESI mTABLET. PAD was defined as an ABPI Results A total of 102 limbs from 62 patients were assessed. Device failure occurred in 20 limbs, resulting in a total of 82 limbs for analysis. MESI-derived ABPI and TBPI showed strong correlation with manual measurements (ρ = 0.850, P Conclusion The MESI mTABLET demonstrates favourable diagnostic performance and substantial time-saving in this cohort. However, it overestimates ABPI/TBPI values and alternative cutoffs may be necessary to maximize sensitivity and specificity.

在知镜精读原文来源DOI 原文

知镜提供医学文献检索、阅读笔记、人体图谱、统计绘图与研究复现工具。