Radiation Dose Optimization in Coronary CT Angiography: A Quality Improvement Initiative.
文献摘要
Background Coronary computed tomography angiography (CCTA) is increasingly used for the evaluation of coronary artery disease; however, radiation exposure remains a patient safety concern. Data on radiation dose optimization from low- and middle-income countries are limited. Project rationale Tabba Heart Institute Karachi identified opportunities to reduce CCTA radiation exposure through implementation of heart rate and body mass index-guided acquisition protocols while maintaining image quality. Project summary A quality improvement initiative was implemented using patient-specific scanning protocols. A total of 422 CCTA examinations were evaluated, including 117 preintervention and 305 postintervention studies. Median effective radiation dose decreased by 45.5%, from 8.8 mSv to 4.8 mSv (P Take-home message Technologist-driven protocol optimization can substantially reduce CCTA radiation exposure without compromising image quality and may serve as a scalable model for cardiovascular imaging programs in resource-constrained settings.